Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
Reason for recall: Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test. Distribution: US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet Quantity: 11182 programs Lot/code info: Programmer Model 3650; Software Model 3330; GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)
05414734509725
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.