Merge Healthcare Merge Application Server Software Release and Merge Healthcare Merge Cardio Workstation Software Release - Product Usage: VERICIS is a system intended to be used to acquire, store, print, transfer, and archive clinical information from Camtronics and other vendors systems including
Check against the official notice below — it lists the full affected range.
Merge Healthcare issued a recall of 62 units of Merge Cardio Workstation Software and Merge Healthcare Merge Application Server Software distributed nationwide because of an error in the calculation of Michigan Reference Ranges that can generate inaccurate Z-Score calculations. This error may reduce the sensitivity of the Z-Score in detecting abnormalities, potentially resulting in false negatives. The affected software versions are Merge Cardio versions 10.X and 11.X with Knowledge Base versions 3.3, 3.4, 3.5, or 5.0 installed on or before February 21, 2020, distributed across multiple states including Illinois, Texas, Oregon, Arizona, Ohio, Tennessee, California, Utah, Missouri, Georgia, New Mexico, Oklahoma, Virginia, Indiana, Connecticut, and Vermont.
The recall notice does not specify a remedy for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative.
Reason for recall: An error in the calculation of the Michigan Reference Ranges can generate an inaccurate Z-Score calculation. The error may reduce the sensitivity of the Z-Score in detecting an abnormality resulting in a false negative. Distribution: US Nationwide distribution including in the states of IL, TX, OR, AZ, OH, TN, CA, UT, MO, GA, NM, OK, VA, IN, CT, and VT. Quantity: 62 units Lot/code info: Merge Cardio versions 10.X and 11.X with Knowledge Base versions 3.3, 3.4, 3.5 or 5.0 installed on or before February 21, 2020. Version 10.0 Pre-UDI Application server PN 88-00528-00; Workstation PN:88-00534-00 Version 10.1.2 Pre-UDI Application server PN 88-01075-00; Workstation PN 88-01078-00 Version 10.2 P2 Pre-UDI Application Server PN 88-01203-00 Version 10.2 P1&P2 Pre-UDI Software Release PN 87-00247-00 Version 10.3 UDI: (01) 00842000100041(10)10.3.0.1137(11)170104 Application server PN: 88-01233-00; Workstation PN 88-01239-00 Version 10.3 Patch 1 UDI: (01)00842000100041(10)10.3.0.2009(11)170309 Software Release PN 88-01319-00; Workstation PN: 88-01320-00 Version 10.3.1 UDI: (01) 00842000100041 (10) 10312135 (11) 171024 Application Server PN: 88-01426-00; Workstation PN 88-01430-00 Version 11.0.2 UDI: (01)00842000100416(10)11.0.2.1523(11)180809 Application Server Installation PN: 88-01570-00; Workstation Installation PN 88-01571-00 Version 11.0.3 UDI: (01)00842000100416(10)11.0.3.1601(11)181019 Workstation Installation PN: 88-01589-00 Version 11.1.1 UDI: (01)00842000100751(10)11.1.1(11)190724 Application Server 30-01650-00; Workstation PN:30-01654-00