Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.
Reason for recall: Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results. Distribution: US Nationwide distribution in the states of CA, DC, GA, OH, OK, MN, MO, NJ, VA and Puerto Rico. Quantity: 37 US 194 PR Lot/code info: Catalogue Number/GTIN (1) UR3825 05055273206906; (2) UR3873 05055273206913 ; (3) UR8334 05055273209600; (4) UR8070 05055273209594. Not Batch Specific.
05055273206906, 05055273206913, 05055273209600, 05055273209594