← Search recalls
FDA (Device enforcement)Recalled 2025-07-21class i

Tandem Diabetes Care, Inc.: Battery Operated infusion pumps. t:slim X2 G5, Replacement, Refurbished, REF: 1000911; t:slim X2 GS Classic, Replacement, Refurbished, REF: 1003808; t: slim X2, Basal-IQ, m

Hazard

Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.

Details from the source

Reason for recall: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization. Distribution: Worldwide distribution. US nationwide including Puerto Rico, Canada, Czech Republic, Slovakia, Spain, United Kingdom, Belgium, Luxembourg, Australia, France, Italy, New Zealand, Finland, Sweden, Denmark, Norway, Saudi Arabia, Germany, Netherlands, Ireland, Portugal, South Africa, Switzerland Quantity: 171,750 Lot/code info: REF/UDI-DI: 1000911/t:slim X2 G5, Replacement, Refurbished/00853052007318; 1003808/t:slim X2 GS Classic, Replacement, Refurbished/ 00853052007981; 1004219/t: slim X2, Basal-IQ, mg/dl/ 00853052007998; 1004484/Pump, t:slim X2, Clinical Use Only; 1005611/Pump, t:slim X2, Control-IQ, mmoUL/ 00850006613304/; 1005613/Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement/ 00850006613311/; 1005615/Pump, t:slim X2, Control-IQ, mg/dl/ 00850006613328/; 1005617/Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement/ 00850006613335/; 1006379/Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement/ 00850006613731/; 1006402/Pump, t:slim X2, Control-IQ 7.4/ 00850006613762/; 1006406/Pump, t:slim X2, Control-IQ 7.4, Replacement/ 00850006613786/; 1006408/Pump, t:slim X2, Control-IQ 7.4/ 00850006613793/; 1006419/Pump, t:slim X2, Basal-IQ, mmoUL/ 00850006613809/; 1006420/Pump, t:slim X2, Basal-IQ, mg/dL 00850006613816; 1006535/Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement/ 00850006613823/; 1006537/Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement/ 00850006613830/; 1008255/Pump, t:slim X2, Control-IQ, mg/dl , Replacement/00850018992671; 1008256/Pump, t:slim X2, Control-IQ, mmoUL, Replacement/00850018992664; 1008257/Pump, t:slim X2, Basal-IQ, mmoUL, Replacement/00850018992657; 1008258/Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement/00850018992640; 1010004/Pump, t:slim X2, Basal-IQ/00389152000015; 1010005/Pump, t:slim X2, Basal-IQ, Replacement/00389152000022; 1010006/Pump

UPC codes

00853052007318, 00853052007981, 00853052007998, 00850006613304, 00850006613311, 00850006613328, 00850006613335, 00850006613731, 00850006613762, 00850006613786, 00850006613793, 00850006613809, 00850006613816, 00850006613823, 00850006613830, 00850018992671, 00850018992664, 00850018992657, 00850018992640, 00389152000015, 00389152000022, 00389152000039, 00389152000046, 00389152000077, 00389152000084, 00389152000091, 00389152000107, 00389152286716, 00389152286815, 00389152286914, 00389152319513, 00389152319612, 00389152319711, 00389152319810

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify