Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization.
Reason for recall: Insulin pumps manufactured with speakers from Revision A and B may malfunction ("Malfunction 16") which prevents audible alerts and causes the pumps to become inoperable and stop infusing insulin which could result in a hyperglycemic event, which may lead to injury and/or hospitalization. Distribution: Worldwide distribution. US nationwide including Puerto Rico, Canada, Czech Republic, Slovakia, Spain, United Kingdom, Belgium, Luxembourg, Australia, France, Italy, New Zealand, Finland, Sweden, Denmark, Norway, Saudi Arabia, Germany, Netherlands, Ireland, Portugal, South Africa, Switzerland Quantity: 171,750 Lot/code info: REF/UDI-DI: 1000911/t:slim X2 G5, Replacement, Refurbished/00853052007318; 1003808/t:slim X2 GS Classic, Replacement, Refurbished/ 00853052007981; 1004219/t: slim X2, Basal-IQ, mg/dl/ 00853052007998; 1004484/Pump, t:slim X2, Clinical Use Only; 1005611/Pump, t:slim X2, Control-IQ, mmoUL/ 00850006613304/; 1005613/Pump, t:slim X2, Control-IQ, mmoUL, Refurbished Replacement/ 00850006613311/; 1005615/Pump, t:slim X2, Control-IQ, mg/dl/ 00850006613328/; 1005617/Pump, t:slim X2, Control-IQ, mg/dl, Refurbished Replacement/ 00850006613335/; 1006379/Pump, t:slim X2, Basal-IQ 6.4, Refurbished Replacement/ 00850006613731/; 1006402/Pump, t:slim X2, Control-IQ 7.4/ 00850006613762/; 1006406/Pump, t:slim X2, Control-IQ 7.4, Replacement/ 00850006613786/; 1006408/Pump, t:slim X2, Control-IQ 7.4/ 00850006613793/; 1006419/Pump, t:slim X2, Basal-IQ, mmoUL/ 00850006613809/; 1006420/Pump, t:slim X2, Basal-IQ, mg/dL 00850006613816; 1006535/Pump, t:slim X2, Basal-IQ, mmoUL, Refurbished Replacement/ 00850006613823/; 1006537/Pump, t:slim X2, Basal-IQ, mg/dL, Refurbished Replacement/ 00850006613830/; 1008255/Pump, t:slim X2, Control-IQ, mg/dl , Replacement/00850018992671; 1008256/Pump, t:slim X2, Control-IQ, mmoUL, Replacement/00850018992664; 1008257/Pump, t:slim X2, Basal-IQ, mmoUL, Replacement/00850018992657; 1008258/Pump, t:slim X2, Basal-IQ, mg/dl ,Replacement/00850018992640; 1010004/Pump, t:slim X2, Basal-IQ/00389152000015; 1010005/Pump, t:slim X2, Basal-IQ, Replacement/00389152000022; 1010006/Pump
00853052007318, 00853052007981, 00853052007998, 00850006613304, 00850006613311, 00850006613328, 00850006613335, 00850006613731, 00850006613762, 00850006613786, 00850006613793, 00850006613809, 00850006613816, 00850006613823, 00850006613830, 00850018992671, 00850018992664, 00850018992657, 00850018992640, 00389152000015, 00389152000022, 00389152000039, 00389152000046, 00389152000077, 00389152000084, 00389152000091, 00389152000107, 00389152286716, 00389152286815, 00389152286914, 00389152319513, 00389152319612, 00389152319711, 00389152319810
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.