Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.
Reason for recall: Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays. Distribution: Worldwide - US Nationwide and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belarus, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Macao, Malaysia, Mexico, Panama, Paraguay, Peru Quantity: 41,824 units Lot/code info: Part Number: 0020300200; UDI-DI: 08426950472490; Lot Numbers: N0330228, N0632773, N0935135, N1036592, N0148986, N0441476, N0643673, N0845161, N1147397;