GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
Reason for recall: GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user. Distribution: US: IL, MA, NM, NY, OK, Quantity: 1 unit Lot/code info: a) H45041SW, UDI/DI *, Serial Numbers: 001423S70