← Search recalls
FDA (Device enforcement)Recalled 2024-05-02class ii

GE Vingmed Ultrasound As: GE Vivid Ultrasound products, Model Numbers: a) Vivid S70, H45041SW

Hazard

GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.

Details from the source

Reason for recall: GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user. Distribution: US: IL, MA, NM, NY, OK, Quantity: 1 unit Lot/code info: a) H45041SW, UDI/DI *, Serial Numbers: 001423S70

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.