GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user.
Reason for recall: GE HealthCare has become aware of an issue with certain Vivid ultrasound systems where, due to a slow DICOM Worklist response time, incorrect patient name and demographics can be displayed on the ultrasound scanner screen rather than the intended patient selected by the user. Distribution: US: IL, MA, NM, NY, OK, Quantity: 7 units Lot/code info: a) H45581LB, UDI/DU *, Serial Numbers: AU01170, AU01180, AU50159; b) H45581D, UDI/DI *, Serial numbers: AU01095, AU50476, AU50252, AU01002