High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
Reason for recall: High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient. Distribution: US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA. Quantity: 858 Lot/code info: REF/UDI-DI/Lot(Expiration): K10-00353K/00884450179246/H3127803(1/4/2028), H3166061(2/24/2028), H3172299(3/14/2028); K10-01603C/00884450593639/H3163984(3/2/2028), H3176777(3/23/2028); K10-01694Q/00884450239308/H3116890(10/31/2027), H3152493(2/3/2028), H3177622(3/14/2028); K10-05159B/00884450153598/H3116922(12/23/2027), H3119571(1/4/2028), H3127907(1/11/2028), H3152490(2/10/2028), H3152491(2/10/2028), H3152492(2/10/2028); K10-05743/00884450456767/H3122012(1/6/2028), H3128429(1/13/2028), H3177446(3/16/2028); K10-05825AP/00884450808108/H3112585(11/27/2027)
00884450179246, 00884450593639, 00884450239308, 00884450153598, 00884450456767, 00884450808108
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.