← Search recalls
FDA (Device enforcement)Recalled 2023-06-12class ii

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion 7 M20. X-ray Image guided Therapy System.

Hazard

The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage.

Details from the source

Reason for recall: The bolts supporting the FlexMove Carriage may become loose and/or break, and cracks may appear in the suspension rail of the FlexMove Carriage. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Indonesia, Iran, Ireland, Italy, Japan, Jordan, Kenya, South Korea, Libya, Lithuania, Malaysia, Maldives, Morocco, Netherlands, New Caledonia, New Zealand, Norway, Oman, Panama, Philippines, Po Quantity: 648 units Lot/code info: Model No.: 722079, 722224; UDI-DI: (01)00884838085268(21), (01)00884838099258(21); System Equipment Number: 69991047, 69991048, 71927518, 71929379, 71929381, 71929386, 71934809, 71935340, 71937626, 71939045, 71957323, 72045694, 72150041, 72173543, 72273457, 72312081, 72321356, 72327897, 72400707, 72414989, 72419144, 72426402, 72648857, 72719305, 72719820, 72721959, 72748229, 72771284, 72805530, 72933546, 73000862, 73003437, 73011242, 73051857, 73337310, 73337321, 73339454, 73340554, 73407813, 73461447, 73516696, 73581917, 73581968, 73616732, 73625357, 73633104, 73717291, 73861967, 74017556, 74020682, 74095702, 74154690, 74154691, 74155276, 74169049, 74227230, 74227231, 74238084, 74250404, 74288937, 74298715, 74369399, 74465159, 74517090, 74582619, 74592051, 74593412, 74608935, 74609935, 74610486, 74610632, 74611953, 74613538, 74620302, 74620341, 74697609, 74817549, 74879566, 74896878, 75215416, 75387595, 75390578, 75434973, 75551701, 75581576, 75610667, 75635433, 75714884, 75844170, 75852749, 75855479, 75935287, 76078970, 76091821, 76138290, 76417103, 76479323, 76515054, 76553358, 76577152, 76787680, 76891085, 77121963, 77427180, 77530081, 77654726, 77654982, 77777427, 77779946, 77970791, 77970948, 77992050, 78105115, 78248928, 78294148, 78301055, 78397953, 78566369, 78742461, 78871714, 78951530, 79075848, 79092310, 79109133, 79112403, 79170704, 79221166, 79260136, 79386139, 79721725, 79755150, 79871193, 80031529, 80162299, 80389257, 80518631, 80634403, 80757669, 80864517, 80

UPC codes

00884838085268, 00884838099258

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify