Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration.
Reason for recall: Cell Collection Device may be at increased risk of the sponge detaching from the string during removal of the device from the patient, which could lead to device fragments in patient, obstruction, airway obstruction, secondary intervention, secondary intervention (with the primary procedure), supraglottic airway obstruction, and aspiration. Distribution: Worldwide - US Nationwide distribution in the states of NY, MN, VA, WI and the countries of Canada, Netherlands, Poland, United Kingdom. Quantity: 6902 Lot/code info: REF/UDI-DI/Lot: CYTO-201/20884521767789/22E1619JZX, 22F0680JZX, 22F0950JZX, 22G0007JZX, 22G0008JZX, 22G0723JZX, 22G0724JZX, 22H0017JZX, 22H0018JZX, 22H0019JZX, 22H0820JZX, 22H0821JZX, 22H1158JZX, 22H1159JZX, 22H1161JZX, 22J0162JZX, 22J0163JZX, 22J0164JZX, 22J1030JZX, 22K0966JZX, 22K0967JZX, 23B0301JZX, 23B0303JZX; CYTO-KIT-USM/10884521546257/22G0635JZX, 22E1624JZX, 22E1625JZX.