← Search recalls
FDA (Device enforcement)Recalled 2024-05-13class ii

CareFusion 303, Inc.: BD Pyxis, Name/REF: BD PYXIS MB TOWER MAIN/ 169-153, BD PYXIS MB TOWER MAIN CR/ 169-154, BD PYXIS MB TOWER MAIN W/MP/ 169-157, BD PYXIS MB TWR MN 1FM-10HH-1FH MP/ 169-143, BD PYX

Hazard

Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment.

Details from the source

Reason for recall: Multiple socket-outlet power strips that shipped with automated dispensing cabinet devices have not been certified to International Electrotechnical Commission (IEC) electrical standards when used as a system with automated dispensing cabinet devices, so use may expose user to leakage currents that could cause varying degrees of harm from a minor shock to permanent impairment. Distribution: US Nationwide distribution including in the states of NY, CA, MO, AZ, CO, KS, IL, FL, OK, OH, TN, GA, NM, NV, NE, TX, NC, KY, WI, SC, IA, MT, SD, VA, NJ, PA, AK, MS, MI, WA, UT, ND, MA, MN, OR, IN, WV, AL, ID, LA, CT, WY, MD, NH, AR, HI, DC. Quantity: 3002 Lot/code info: UDI-DI: 10885403512568. Devices Manufactured November 15, 2012 - July, 25 2023 User Guide, DIR 10000414989/DIR 10000414987/DIR 10000414990, Version 00.

UPC codes

10885403512568, 10000414989, 10000414987, 10000414990

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify