FDA (Device enforcement)Recalled 2023-07-12class ii

Siemens Medical Solutions USA, Inc.: The following Pro.specta Emission Computed Tomography Systems that were manufactured with an SLD Look Ahead Optical Sensor Assembly: (a) Symbia Pro.specta Q3, Mode

Hazard

During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution.

Details from the source

Reason for recall: During a whole-body planar scan with auto-contour, there is a possibility the short-linear drive (SLD) look-ahead sensors may trigger earlier than expected and may result in detector 1 positioning further from the patient than needed leading to a reduction in image resolution. Distribution: Distribution was made to ID, KS, MA, NH, OH, TX, and VA. There was no government/military distribution. Foreign distribution was made to Argentina, Australia, Austria, Belgium, Canada, Colombia, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, Netherlands, Portugal, Singapore, Switzerland, Turkey, United Arab Emirates, and United Kingdom. Quantity: 108 systems Lot/code info: Software version VA20A. (a) Serial numbers 100201, 100202, 100203, 100204, 100205, 100206, 100207, 100209, 100208, 100210, 100212, 100211, 100213, 100215, 100217, 100219, 100218, 100220, 100221, and 100224; UDI-DI 4056869264745. Additional serial numbers as of 09/07/2023: 100223, 100225, 100227, 100228, 100237; Additional serial numbers as of 10/17/2023: 100214, 100229; Additional serial numbers as of 12/12/2023: 100232, 100231, 100234; Additional serial number as of 01/15/2024:100226. (b) Serial numbers 100101, 100102, 100201, 100203, 100204, 100207, 100206, 100208, 100205, 100209, 100210, 100212, 100213, 100214, 100215, 100216, 100218, 100217, 100219, 100220, 100221, 100223, 100224, 100222, 100226, 100228, 100227, 100225, 100229, 100231, 100232, 100233, 100235, 100234, 100237, 100239, 100241, 100240, 100236, 100238, 100242, 100243, 100247, 100244, 100250, 100251, 100252, and 100255; UDI-DI 4056869264752; Additional serial numbers as of 09/07/2023: 100245, 100256, 100257, 100258, 100261, 100262, 100265, 100273; Additional serial numbers as of 10/17/2023: 100230, 100266; Additional serial numbers as of 12/12/2023: 100269; Additional serial numbers as of 01/15/2024: 100248, 100254, 100260, 100271, 100272. (c) Serial numbers 100101, 100201, 100203, 100202, 100205, and 100206; UDI-DI 4056869264769; Additional serial numbers as of 09/07/2023: 100208, 100210; Additional serial numbers as of 12/12/2023: 100209

UPC codes

4056869264745, 4056869264752, 4056869264769

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify