← Search recalls
FDA (Device enforcement)Recalled 2023-06-27class ii

Zimmer, Inc.: Zimmer Periarticular Locking Plate (ZPLP) Distal Lateral Femoral Plate 6 holes 159 mm Length-Indicated for temporary internal fixation and stabilization of osteotomies and fractures, inc

Hazard

Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.

Details from the source

Reason for recall: Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock. Distribution: AR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan Quantity: 75 units Lot/code info: Lot Number/UDI: 64254511 /(01)0089024055940(10)64254511; 64296571/(01)0089024055940(10)64296571; 65379087/(01)0089024055940(10)65379087; 65379093/(01)0089024055940(10)65379093; 65379096/(01)0089024055940(10)65379096

UPC codes

0089024055940

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify