← Search recalls
FDA (Device enforcement)Recalled 2023-05-30class ii

Johnson & Johnson Surgical Vision, Inc.: VERITAS Advanced Infusion Packs (VRT-AI) and VERITAS Advanced Fluidics Packs (VRT-AF)

Hazard

Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.

Details from the source

Reason for recall: Weld protrusion/physical gap between the housing and pack cover, may exceed design specification, due to a manufacturing issue, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema. Distribution: Worldwide - US Nationwide distribution including in the states of NC, OH, NY, IN, FL, TN, CA, PA, AZ, AL, TX, WA, WV, MS, GA, UT, KY, IA, MO, AR, DE, IL, WI, VA, NJ, LA, SC, HI, OR, VT, MN, NV, CO and the countries of Argentina, Australia, Austria, Azerbaijan, Belgium, Brazil, Bulgaria, Canada, Chile, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, In Quantity: 148,941 Lot/code info: All lots within their expiration date as of 5/29/2023. Part Number/UDI-DI/Lot: VRT-AI/05050474700901/60298931, 60302489, 60302939, 60302941, 60302953, 60302954, 60304170, 60304175, 60304176, 60305660, 60305661, 60305662, 60305665, 60306934, 60306935, 60306936, 60306940, 60306942, 60308193, 60309850, 60309844, 60309845, 60309853, 60314642, 60314676, 60314677, 60316112, 60351692, 60372490, 60372491, 60374175, 60374945, 60376930, 60378840, 60378842, 60378844, 60400795, 60401984, 60413115, 60413116, 60413117, 60414191, 60415077, 60415078, 60416304, 60420003, 60421952, 60425389, 60426314, 60426315, 60428217, 60428216, 60429442, 60433653, 60433654, 60433655, 60435921, 60435922, 60435923, 60437988, 60437989, 60440239, 60440240, 60442450, 60442451, 60442452, 60442453, 60442454, 60444535, 60446059, 60448847, 60448856, 60448857, 60376931, 60379976, 60429443. VRT-AF/05050474700895/60304172, 60304173, 60304174, 60305657, 60305658, 60305659, 60306930, 60306931, 60306932, 60306933, 60308189, 60308190, 60309848, 60309847, 60314640, 60314641, 60314674, 60314675, 60378839, 60420004, 60446054, 60446055

UPC codes

05050474700901, 05050474700895

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify