GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Reason for recall: GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station). Distribution: Worldwide distribution. Quantity: 2366 units Lot/code info: No UDI, Serial Number or Sales Order Number: KSA7536146, KSA7536158, KSA7536159, KSA7550156, KSA7550158, KSA7550159, KSA7667317, KSA7754389, KSA7769597, KSA7769610, KSA7769612, KSA7773420, KSA7773431, KSA7773440, KSA7120704, KSA7120767, KSA6964886, KSA6808842, KSA6808852, KSA6808866, KSA6808868, KSA6808881, KSA6808883, KSA6808912, KSA6808928, KSA6866974, KSA6664007, KSA6866963, KSA7120766, KSA7162586, KSA6623197, KSA6623216, KSA6623223, KSA6623226, KSA6623239, KSA6623248, KSA6623270, KSA6623278, KSA7391460, KSA6477094, KSA6477095, KSA6477135, KSA6477155, KSA6477159, KSA6477206, KSA6954245, KSA6954269, KSA6954276, KSA6954277, KSA6954284, KSA6954291, KSA6477096, KSA6477154, KSA7702196, KSA7702198, KSA7702199, KSA7702208, KSA7702222, KSA6623191, KSA6623194, KSA6623237, KSA6623251, KSA6623267, KSA6626626, KSA6626632, KSA6626639, KSA6626666, KSA6626670, KSA6626716, KSA6536442, KSA6536443, KSA6536454, KSA6536460, KSA6536469, KSA6536484, KSA6536485, KSA6536490, KSA6536496, KSA6536504, KSA6536526, KSA6536528, KSA6536532, KSA7120699, KSA7120728, KSA7120764, KSA7120773, KSA7162570, KSA7162593, KSA7162610, KSA7162622, KSA7162640, KSA6623198, KSA6623202, KSA6623232, KSA6623236, KSA6623275, KSA6623280, KSA6626622, KSA6626625, KSA6626635, KSA6626636, KSA6626648, KSA6626655, KSA6626659, KSA6626661, KSA6536520, KSA6536525, KSA7304431, KSA6866921, KSA6866926, KSA6866927, KSA6866983, KSA6808882, KSA6649195, KSA6649250, KSA7162575, KSA7162588, KSA7162608, KSA7162614, KSA7162656, KSA6899015, KSA
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.