Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4
Check against the official notice below — it lists the full affected range.
Abiomed, Inc. is recalling Automated Impella Controller (AIC) consoles distributed worldwide, including in the US and multiple other countries. The recall affects 625 units across four product codes due to two identified issues: certain AIC consoles experienced alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing; additionally, improper alignment between the purge cassette and motor drive could cause a piston block event that prevents the purge system from priming properly, requiring users to switch to a backup AIC console.
The specific serial numbers affected are listed by product code and region. The recall record does not specify a remedy for affected consumers.
Written by software from the official notice below. Not reviewed by a lawyer.
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
Reason for recall: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Denmark, France, Germany, Italy, Kuwait, Netherlands, Norway, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom. Quantity: 625 units Lot/code info: 1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595, IC1598, IC1664, IC1665, IC1715. 2. Product Code: 0042-0000-EU. GTIN: 00813502011289. Serial Numbers: IC1217, IC1265, IC1275, IC1276, IC1288, IC1284, IC1301, IC1302, IC1304, IC1305, IC1344, IC1341, IC1343, IC1340, IC1345, IC1350, IC1347, IC1349, IC1383, IC1381, IC1384, IC1386, IC1393, IC1396, IC1397, IC1395, IC1413, IC1415, IC1411, IC1416, IC1490, IC1488, IC1522, IC1523, IC1524, IC1525, IC1547, IC1549, IC1546, IC1548, IC1578, IC1577, IC1579, IC1580, IC1594, IC1584, IC1590, IC1592, IC1591, IC1585, IC1588, IC1653, IC1676, IC1685, IC1684, IC1677, IC1678, IC1699, IC1700. 3. Product Code: 0042-0000-UK. GTIN: 00813502011296. Serial Numbers: IC1387, IC1389, IC1388, IC1519, IC1542, IC1545, IC1581, IC1582, IC1583, IC1573. 4. Product Code: 0042-0000-US. GTIN: 00813502010022. Serial Numbers: IC1004, IC1002, IC1001, IC1006, IC1007, IC1008, IC1014, IC1015, IC1017, IC1019, IC1024, IC1025, IC1028, IC1029, IC1026, IC1027, IC1033, IC1035, IC1031, IC1032, IC1036, IC1037, IC1038, IC1039, IC1041, IC1042, IC1044, IC1049, IC1043, IC1046, IC1053, IC1055, IC1052, IC1040, IC1030, IC1056, IC1047, IC1051, IC1050, IC1060, IC1061, IC1062, IC1064, IC1065, IC1066, IC1067, IC1074, IC1070, IC1071, IC1072, IC1073, IC1068, IC1069, IC1058, IC1075, IC1076, IC1077, IC1078, IC1079, IC1084, IC1085, IC1083, IC1080, IC1088, IC1086, IC1087, IC1081, IC1082, IC1100, IC1099, IC1093, IC1095,