Double Offset Rasp Handle, Left Hand A/S Rasp Handle (Left)
Check against the official notice below — it lists the full affected range.
Zimmer, Inc. is recalling 412 units of Double Offset Rasp Handle, Left Hand A/S Rasp Handle (Left) due to the handle failing to remain assembled to the rasp during use. The company received complaints about various failure modes, including fracture of the leaf spring, which could cause the handle to detach from the rasp while being used.
The recalled product was distributed worldwide to the United States (including Alaska, Alabama, Arizona, California, Colorado, Florida, Georgia, Indiana, Kansas, Kentucky, Michigan, Minnesota, North Carolina, New Jersey, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, and Virginia) and to Australia, France, Germany, Italy, Switzerland, and Canada. The affected item numbers and lot numbers are Item Number 00780803521 with Lot Numbers 56474891, 56474892, 56474893, and 56474894. The source does not specify a remedy or what affected consumers should do.
Written by software from the official notice below. Not reviewed by a lawyer.
The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring.
Reason for recall: The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring. Distribution: Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, IN, KS, KY, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, and VA, and the countries of Australia, France, Germany, Italy, Switzerland, and Canada. Quantity: 412 units Lot/code info: Item Number 00780803521 Lot Number 56474891, 56474892, 56474893, 56474894
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.