Vl DBL OFFST WI LRG STPL, RT Nonsterile
Check against the official notice below — it lists the full affected range.
Zimmer, Inc. has recalled 412 units of its VL DBL OFFST WI LRG STPL, RT Nonsterile rasp due to a hazard in which the handle may fail to remain assembled to the rasp during use. The company received complaints about the handle detaching due to various failure modes, including fracture of the leaf spring. The affected product has item number 82016757942 and lot numbers 97009220, 97009255, 97009256, and 97009585.
The recalled rasps were distributed worldwide, including throughout the United States in 24 states and in Australia, France, Germany, Italy, Switzerland, and Canada. The recall source does not specify a remedy or what affected consumers should do regarding this product.
Written by software from the official notice below. Not reviewed by a lawyer.
The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring.
Reason for recall: The firm received complaints for failure of the handle to remain assembled to the rasp during use due to various failure modes, including fracture of the leaf spring. Distribution: Worldwide Distribution-USA (nationwide) including the states of AK, AL, AZ, CA, CO, FL, GA, IN, KS, KY, MI, MN, NC, NJ, NY, OH, OK, OR, PA, SD, TN, TX, UT, and VA, and the countries of Australia, France, Germany, Italy, Switzerland, and Canada. Quantity: 412 units Lot/code info: Item Number 82016757942 Lot Number 97009220, 97009255, 97009256, 97009585
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.