← Search recalls
FDA (Device enforcement)Recalled 2023-06-23class ii

Datascope Corp.: TRANS-RAY IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number 7FR. 34cc TRANS-RAY IAB WITH ACCES

Hazard

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

Details from the source

Reason for recall: During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death. Distribution: Domestic distribution nationwide. International distribution worldwide. Quantity: 321,609 total kits Lot/code info: Product Description UDI 7FR. 34cc TRANS-RAY IAB WITH ACCESSORIES 10607567107656 TRANS-RAY 7FR. 40cc IAB WITH ACCESSORIES 10607567107663 TRANS-RAY PLUS 7.5Fr 35cc IAB WITH ACCESSORIES 10607567111711 TRANS-RAY PLUS 7.5Fr 40cc IAB WITH ACCESSORIES 10607567111728 PACKAGED INSERTION KIT - TRANS-RAY 7 Fr. 34/40 cc IABs 10607567107684 PACKAGED INSERTION KIT - TRANS-RAY PLUS 7.5 Fr. 35/40 IABs 10607567111810

UPC codes

10607567107656, 10607567107663, 10607567111711, 10607567111728, 10607567107684, 10607567111810

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.