Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.
Reason for recall: Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system. Distribution: Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAYMAN ISLANDS, CHILE, COLOMBIA, COSTA RICA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ESTONIA, FINLA Quantity: 3,425 Lot/code info: REF/Serial Number or Range: 201-30300/ L00610-0004-L06786-0002 201-90401/ 19234129-20119751 and L00610-0011-L08005-0001 201-90411/19234128-20190463 and L03849-0004-L08002-0001 201-90421/19474336-19606873 and L04780-0005-L07601-0001 201-90701/ 19415572 and L03774-0007-L08007-0001 L201-90401/ L00694-0006-L06579-0004 and L201-90401 L201-90411/ 19258710-19605455 and L01627-0007-L07980-0001 L201-90421/ L01716-0010-L06482-0005 UDI-DI: 07640135140689, 07640135140726, 07640135140702, 07640135140719, 07640135140856, 05415067037282, 07640135140894, 07640135149019, 07640135149002, 07640135140702, 07640135140719
07640135140689, 07640135140726, 07640135140702, 07640135140719, 07640135140856, 05415067037282, 07640135140894, 07640135149019, 07640135149002
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.