FDA (Device enforcement)Recalled 2024-05-17class ii

Siemens Healthcare Diagnostics Inc: Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPoint 500e Blood Gas System (China) 11416752; RAPIDPoint 500e

Hazard

Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions.

Details from the source

Reason for recall: Siemens Healthineers has identified an issue with the RAPIDPoint500e system software version 5.3. This issue affects how the sample source is identified at the LIS when using the capillary mode. With this software version, samples run in the capillary mode are incorrectly labeled as arterial when displayed on the LIS, leading to the potential for results to be flagged according to an established arterial range. Capillary sample results are correctly identified as capillary on both the RAPIDPoint 500e system display and the instrument printouts. The issue only affects the new software version 5.3 and only capillary sample mode is impacted. The inability to distinguish specimen sources for blood gas may lead to incorrect diagnoses and management decisions. Distribution: Worldwide - US Nationwide distribution in the states of FL, GA, LA, MI, NC and the countries of Chile, Costa Rica, France, Germany, Japan, Lebanon, Mexico, Netherlands, New Zealand, Poland, Spain, Turkey, United Kingdom. Quantity: 9 units US Lot/code info: UDI-DI: (1)RAPIDPoint 500e Blood Gas System (USA) 00630414286143; (2)RAPIDPoint 500e Blood Gas System (China) 00630414286150; (3) RAPIDPoint 500e Blood Gas System (Japan) 00630414286167; (4) RAPIDPoint 500e Blood Gas System (ROW) 00630414286174 Software Version 5.3.

UPC codes

00630414286143, 00630414286150, 00630414286167, 00630414286174

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify