Abbott Ireland Diagnostics Division: Alinity i Ferritin Reagent Kit, List Number 07P6530 The Alinity i Ferritin assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative d

Is my product affected?

Alinity i Ferritin Reagent Kit, List Number 07P6530 The Alinity i Ferritin assay is a chemiluminescent microparticle immunoassay (CMIA) used for the quantitative determination of ferritin in human serum and plasma on the Alinity i analyzer.

  • UPC 00380740131487

Check against the official notice below — it lists the full affected range.

What happened

Abbott Ireland Diagnostics Division is recalling Alinity i Ferritin Reagent Kits (List Number 07P6530) due to a manufacturing issue that can damage the bottle neck in some reagent cartridges. The affected lots were distributed to Arkansas, Florida, Georgia, Maryland, Oklahoma, and Virginia. A total of 216 units are involved in this recall. When a bottle neck is damaged, it may cause a failure during the reagent loading process when the cartridge is used with the Alinity i instrument.

The affected lot/code numbers are: UDI/Lot Numbers (01)00380740131487 (17)190730(10)91532UI00/91532UI00 and (01)00380740131487 (17)190905(10)93238UI00/93238UI00. The source document does not specify what consumers should do regarding this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

Details from the source

Reason for recall: Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument. Distribution: US Distribution to states of: AR, FL, GA, MD, OK and VA. Quantity: 216 units Lot/code info: UDI/ Lot Numbers: (01)00380740131487 (17)190730(10)91532UI00/91532UI00, (01)00380740131487 (17)190905(10)93238UI00/93238UI00

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify