Intuitive Surgical, Inc.: da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

Is my product affected?

da Vinci SP Camera; PN 430060-33; Model Number: 430060-33. For endoscopic surgery.

  • UPC 00886874114384

Check against the official notice below — it lists the full affected range.

What happened

Intuitive Surgical, Inc. is recalling 305 da Vinci SP Camera units (Model Number 430060-33) distributed nationwide in the US and South Korea. Some cameras were found to have residual soil stuck to the camera tip after reprocessing, which could reduce image quality and potentially require surgeons to switch to a backup camera or change the surgical approach. If residual soil contacts a patient during surgery, it could cause sensitization that would complicate future blood therapy or tissue typing; additionally, if a small piece of inert epoxy is retained, it could lead to scar tissue formation.

The recall notice does not specify what an affected consumer should do in response to this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form.

Details from the source

Reason for recall: Some cameras were found to have residual soil adhered to the tip of the camera after reprocessing which may affect image quality resulting in needing to switch to a second camera or converting to another surgical modality. If residual soil makes contact with the patient, exposure could result in sensitization making future component blood therapy or tissue typing more challenging; additionally if a small piece of inert epoxy were to be retained, scar tissue could form. Distribution: Distribution US nationwide and South Korea. Quantity: 305 units Lot/code info: Model Number: 430060-33; UDI Number: 00886874114384; All serial numbers.

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify