← Search recalls
FDA (Device enforcement)Recalled 2023-05-24class ii

Stryker Medical Division of Stryker Corporation: Power-PRO 2- A powered ambulance cot that consists of a platform, including a mattress, mounted on a wheeled, retractable X-frame that is designed to s

Hazard

Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.

Details from the source

Reason for recall: Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Switzerland, United Kingdom. Quantity: 1,980 cots Lot/code info: Model: 650700000000 UDI-DI: 07613327504460; Model: 650705550001 UDI-DI: 07613327559118; Model: 650705550002 UDI-DI: 07613327559125; Cots shipped between February 11, 2022 and February 1, 2023 Serial Numbers: 2112001485 2112001488 2112001489 2112001490 2112001491 2112001492 2112001493 2112001494 2112001495 2112001496 2112001501 2112001503 2112001504 2112001508 2112001509 2112001510 2112001511 2112001512 2112001513 2112001514 2112001515 2112001516 2112001517 2112001518 2112001519 2112001520 2112001521 2112001522 2112001523 2112001524 2112001525 2112001526 2112001527 2112001528 2112001529 2112001530 2112001531 2112001532 2112001533 2112001534 2112001535 2112001536 2202000473 2202000474 2202000475 2202000476 2202000477 2202000478 2202000479 2202000480 2202000481 2202000482 2202000483 2202000484 2202000485 2202000487 2202000488 2202000489 2202000490 2202000491 2202000492 2202000797 2202000798 2202000799 2202000800 2202000801 2202000802 2202000803 2202000804 2202000805 2202000806 2202000807 2202000808 2202000809 2202000810 2202000811 2202000812 2202000813 2202000814 2202000815 2202000816 2202000817 2202000819 2202000820 2202000821 2202000822 2202000823 2202000824 2202000825 2202000826 2202001299 2202001300 2202001301 2202001302 2202001303 2202001304 2202001305 2202001306 2202001307 2202001308 2202001309 2202001311 2202001312 2202001313 2202001314 2202001315 2202001316 2202001317 2202001318 2202001518 2202001519 2202001520 2202001521 2202001522 2202001523 2202001524 2202001526 22020

UPC codes

650700000000, 07613327504460, 650705550001, 07613327559118, 650705550002, 07613327559125

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.