Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
Reason for recall: Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection. Distribution: Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA. Quantity: 1,445,600 Lot/code info: REF/UDI-DI/Lot(Expiration): SCXT3-5-2000/10887709097107/14258739(1-Dec-26), 14258743(1-Feb-26), 14258744(1-Dec-26), 14264200(1-Dec-26), 14272241(1-Dec-26), 14272248(1-Dec-26), 14287456(1-Dec-26), 14287459(1-Dec-26), 14287480(1-Dec-26), 14287482(1-Dec-26), 14287485(1-Dec-26), 14294832(1-Dec-26), 14272254(1-Dec-26), 14310313(1-Dec-26), 14310425(1-Dec-26), 14323354(1-Dec-26), 14287475(1-Dec-26), 14294840(1-Dec-26), 14321645(1-Jan-27), 14321652(1-Jan-27), 14331169(1-Jan-27), 14332773(1-Jan-27), 14341732(1-Oct-26), 14294825(1-Jan-27), 14303209(1-Jan-27), 14310308(1-Jan-27), 14355404(1-Jan-27), 14355409(1-Jan-27) SCXT3-2000/20887709079162/14258214(1-Jan-27), 14258206(1-Jan-27), 14258208(1-Feb-27).
10887709097107, 20887709079162
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.