Guardian Connect App CSS7200 iOS and Guardian Connect Transmitter GST4C used on the iPhone, iPad and iPod Touch devices.
Check against the official notice below — it lists the full affected range.
Medtronic Inc. has recalled its Guardian Connect App CSS7200 and Guardian Connect Transmitter GST4C used on iPhone, iPad, and iPod Touch devices. The recall affects approximately 23,978 units distributed nationwide. Customers using these devices with iOS software version 12, 12.1, or 12.2 may experience significantly shortened transmitter battery life, getting approximately 4-5 days of use instead of the normal 7 days or more after a full charge.
The record does not specify a remedy or what affected consumers should do.
Written by software from the official notice below. Not reviewed by a lawyer.
Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge.
Reason for recall: Customers using the firm's continuous glucose monitoring system application on an iPhone, iPad or iPod Touch with ios software version 12, 12.1, or 12,2 are likely to experience a shortened transmitter battery life (approximately 4-5 days instead of the normal 7 days or more) after a full charge. Distribution: US Nationwide distribution. Quantity: 23,978 Lot/code info: Model Number CSS7200 (Guardian Connect App: iOS); Catalog Numbers: MMT-7821 (Guardian Connect Transmitter - GST4C); UPN - 00763000186333. There are no assigned lot or serial numbers.
Official record: FDA (Device enforcement) ↗