GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH Nuclear Medicine System

Hazard

Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury.

Details from the source

Reason for recall: Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector motion may result in life-threatening injury. Distribution: Worldwide - US Nationwide distribution. Quantity: 21 Lot/code info: Serial Numbers: Millennium VG 1074; "Varicam Millennium VG Discovery VH" 1003 1092 10147 10242 10374 10378 10539 10543 10550 10553 10561 10569 10570 10571 10579 10582 10587 10599 10604 3101-1502

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.