Siemens Healthcare Diagnostics, Inc.: Aptio Automation Storage and Retrieval Module (SRM)

Hazard

A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B.

Details from the source

Reason for recall: A firmware error may lead to an incorrect association of test result(s) to a sample ID and therefore to an incorrect patient result. The mis-association of test results, or a delay in testing, may occur when all the conditions listed below occur within a few milliseconds timeframe of each other: - The SRM module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the SRM module buffer - There is divert gate malfunction and a narrow timing window Only in this specific scenario, Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without a unique error message being generated. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the pending test orders (not yet performed) on Tube B. Distribution: Worldwide distribution - US Nationwide distribution in the states of CA, CO, FL, IL, KY, LA, MA, MD, MI, MS, MT, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, & WV. The countries of Australia, Brazil, Canada, China, France, Germany, Greece, Italy, Portugal, Saudi Arabia, Spain, Sweden, & United Kingdom. Quantity: 83 units Lot/code info: Model No. 10713762 and 10715227; UDI-DI: 00630414596754; All units with software versions prior to 38.

UPC codes

00630414596754

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify