enGen (TM) Laboratory Automation Systems using all TCAutomation(TM) Software Versions with the InOut Communication Interface, IVD
Check against the official notice below — it lists the full affected range.
Ortho-Clinical Diagnostics recalled enGen Laboratory Automation Systems using TCAutomation Software due to a confirmed software anomaly that may cause sample IDs to become mismatched with test results. The issue occurs when four specific conditions happen simultaneously: a Bypass module is positioning a sample, at least two additional samples are queued, an RFID tag read failure occurs on the first sample carrier, and the VITROS analyzer is unavailable. When this happens, the software queue becomes out of sync with the physical sample queue, potentially causing test results to be reported under wrong sample IDs, and all subsequent samples would have mismatched results until the Bypass module is reset or the queue empties. Although no patient harm has been reported to date, this could potentially lead to misdiagnosis or inappropriate medical intervention depending on the test type and patient condition.
The recall affects 33 units in the U.S. and 43 units internationally, distributed across multiple states and countries including Canada, Mexico, Brazil, and various European nations. The source document does not specify a remedy or instructions for affected consumers to follow.
Written by software from the official notice below. Not reviewed by a lawyer.
Software anomaly; Thermo-Fisher Scientific initially discovered and Ortho-Clinical Diagnostics, subsequently, confirmed a software anomaly that may potentially result in miss-associated sample IDs involving the communication between enGen TCAutomation (TCA) Bypass modules (manufactured by Thermo-Fisher Scientific) and the VITROS Analyzers. To date, no occurrences of this issue have been observed on Orthos enGen(TM) Systems. Sample IDs may become miss-associated with the actual sample being processed due to discrepant information passed to the analyzer through InOut communication device, triggered by a specific set of events occurring between the Bypass module and the analyzer. --- All of the following conditions must be present for the issue to potentially occur: (1.) The Bypass module is in the process of positioning Sample A at the aspiration position, and (2.) At least two additional sample tubes (Samples B and C) are waiting to be aspirated in the Bypass module queue, and (3.) A Radio Frequency Identification (RFID) Tag Read failure occurs on the Sample A carrier, and (4.) The VITROS System is not available to aspirate on the track (for example, the analyzer is initializing, or the user is making entries on the diagnostic screen, or samples are being aspirated from the front of the analyzer, etc.). --- If the software anomaly occurs and the issue is present, the Sample ID in the INOUT communication device software queue becomes asynchronous with the actual Sample ID physical queue, and the VITROS assay result will be reported for the wrong Sample ID. All subsequent samples will have miss-associated results until the Bypass module is reset or the queue is empty. --- This issue could potentially miss-associate a sample/patient ID with a test result of a different sample. Depending on the test involved, patients health condition, etc., the potential health risk posed to the affected patient could range from no risk to serious risk, and potentially could lead to misdiagnosis or inappropriate medical intervention. To date, no patient harm was reported related to this product issue.
Reason for recall: Software anomaly; Thermo-Fisher Scientific initially discovered and Ortho-Clinical Diagnostics, subsequently, confirmed a software anomaly that may potentially result in miss-associated sample IDs involving the communication between enGen TCAutomation (TCA) Bypass modules (manufactured by Thermo-Fisher Scientific) and the VITROS Analyzers. To date, no occurrences of this issue have been observed on Orthos enGen(TM) Systems. Sample IDs may become miss-associated with the actual sample being processed due to discrepant information passed to the analyzer through InOut communication device, triggered by a specific set of events occurring between the Bypass module and the analyzer. --- All of the following conditions must be present for the issue to potentially occur: (1.) The Bypass module is in the process of positioning Sample A at the aspiration position, and (2.) At least two additional sample tubes (Samples B and C) are waiting to be aspirated in the Bypass module queue, and (3.) A Radio Frequency Identification (RFID) Tag Read failure occurs on the Sample A carrier, and (4.) The VITROS System is not available to aspirate on the track (for example, the analyzer is initializing, or the user is making entries on the diagnostic screen, or samples are being aspirated from the front of the analyzer, etc.). --- If the software anomaly occurs and the issue is present, the Sample ID in the INOUT communication device software queue becomes asynchronous with the actual Sample ID physical queue, and the VITROS assay result will be reported for the wrong Sample ID. All subsequent samples will have miss-associated results until the Bypass module is reset or the queue is empty. --- This issue could potentially miss-associate a sample/patient ID with a test result of a different sample. Depending on the test involved, patients health condition, etc., the potential health risk posed to the affected patient could range from no risk to serious risk, and potentially could lead to misdiagnosis or inappropriate medical intervention. To date, no patient harm was reported related to this product issue. Distribution: MA, NJ, NY, NY, PA,DE, MD, WV, NC, GA, FL, TN, MS, OH, MI, IL, MO, NE, AR, TX, CO, ID, TX, CA and Foreign distribution to the following countries: Belgium, Bermuda, Brazil, Canada, Chile, Denmark, England, France, Italy, Mexico, Norway, Portugal, Romania, Singapore, Spain, Sweden, Taiwan, United Kingdom. Quantity: US: 33 units; OUS: 43 units Lot/code info: US Serial Numbers/J Numbers: B3ZHP2J/JB3ZHP2J; 736BM4J/J736BM4J; 953289/J953289; 953252/J953252; HJ5K84J/JHJ5K84J; FCM073J/JFCM073J; HPKG25J/JHPKG25J; GWXQ52J/JGWXQ52J; GBX395J/GBX395J; JFBX395J/JFBX395J; FXCBG3J/JFXCBG3J; 3RHC23J/J3RHC23J; JG7BFB5J/JG7BFB5J; 953270/J953270; 953221/J953221; 54FC01J/J54FC01J; 953209/J953209; 5J4B45J/J5J4B45J; 5LQ6Q2J/ J5LQ6Q2J; 2Z3YV1J/J2Z3YV1J; FRKMB2J/JFRKMB2J; 5GV6W2J/5GV6W2J; 7KSC15J/J7KSC15J; 953276/J953276; J2QKG25J/J2QKG25J; DK2BG3J/JDK2BG3J; 953266/J953266; J86PWC5J/ J86PWC5J; JXBQW4J/JJXBQW4J; CXL941J/JCXL941J; J953224/J953224; 36PWC5J/J36PWC5J; 4HX7S4J/J4HX7S4J --- Foreign Serial Numbers/J #s: CRKM82/J31265; 2S5OP0J/J31067; C4Z655J/ J31421; 6VR7K4J/J31401; 90JYS22/J31350; HWJZY4J/JHWJZY4J; J3J4B45J/J3J4B45J; JCT20Q3J/JCT20Q3J; J953234/J953234; J41BCY4J/J41BCY4J; J953244/J953244; J1HX7S4J/ J1HX7S4J; J16XQH4J/J16XQH4J; GWJZY4J/ GWJZY4J; DWXO52J/DWXO52J; 17QP94J/J20022; BGV6W2J/ J20003; H5PWC5J/J271371; 2LSSF5J/J271377; 1K5K84J/ J271264; 8W7WG3J/J85206; 96XQHAJ/ J7023645; 1LSSF5J/J7025050; 86T205J/ J7024893; C3JTP4J/J7023847; D3JTP4J/J7023874; C1JNZ5J/J7023730; 7DRJ44J/J7023728; 7YCF43J/J7023435; 18BFB5J/J7025017; J953222/J953222; J2V6LW4J/J2V6LW4J; J56T205J/J56T205J; BGV6W2J/J57663; 4drj44j/J800103; 6KSC15J/J1801116; FWGFC4J/J75663; 7GV6W2J/J75615; BT9PB5J/ J75666; JT6LW4J/ J30372; 38769917923/JTAINAN1; F2250767/J3055503; F1150698/ J3055496.
Official record: FDA (Device enforcement) ↗