← Search recalls
FDA (Device enforcement)Recalled 2026-03-31class ii

Erbe Medical, LLC: ERBEFLO CleverCap: Hybrid Tubing/Cap Set for Olympus Scopes, 20325-201; ERBEFLO CleverCap Hybrid Tubing/Cap Set for Fujifilm Scopes, REF: 20325-203;

Hazard

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

Details from the source

Reason for recall: Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure. Distribution: US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV. Quantity: 23,359 Lot/code info: REF/UDI-DI/Lots: 20325-201/04065655000603/W4466253, W4466254, W4466256, W4466257, W4466258, WO471080, WO471081, WO471083, WO471084, WO473975, WO473976, WO477586, WO477587, WO477588, WO477589, WO477590; 20325-203/04065655001211/WO473986

UPC codes

04065655000603, 04065655001211

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.