← Search recalls
FDA (Device enforcement)Recalled 2021-06-07closedclass ii

BioFire Diagnostics, LLC: FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

Is my product affected?

FilmArray Pneumonia Panel (Pneumo) Part Numbers: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

  • UPC 00815381020178
  • UPC 00815381020185

Check against the official notice below — it lists the full affected range.

What happened

BioFire Diagnostics, LLC is recalling FilmArray Pneumonia Panel test kits (Part Nos. RFIT-ASY-0144 and RFIT-ASY-0145) due to real-time stability study failures that resulted in an increased rate of false negative results between the 6 to 12 month expiry time-frame. A total of 4,462 kits were distributed globally, including throughout the United States, U.S. territories, and numerous countries internationally.

The recall affects test kits with the specified part numbers and UDI codes. The source document does not specify a remedy or recommended action for affected consumers.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame.

Details from the source

Reason for recall: Due to real-time stability study failures resulting in increase rate of false negative results between the 6 to 12 month expiry time-frame. Distribution: U.S.: AK, AL, AZ, CA, CA, CO, CT, DC, DE, FL, GA, Guam, HI, IA, ID, IL, IN, IN, KS, KY, KY, LA, MA, MD, ME, MI, MI, MN, MO, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV and WY. O.U.S.: Algeria, Angola, Anguilla, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bermuda, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Chile, China, Colombi Quantity: 4,462 kits (Globally) Lot/code info: Part No: RFIT-ASY-0144 / UDI: 00815381020178 (30 test kit) Part No: RFIT-ASY-0145 / UDI: 00815381020185 (6 test kit)

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.