Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse.
Reason for recall: Simultaneous ECG measurements with multiple devices, including a manual defibrillator, on the same patient may cause electrical noise or interference, which may lead to distortions of the ECG signal; distortions may lead to no or ineffective pacing, or no or inappropriate defibrillation pulse. Distribution: US Nationwide distribution including in the states of VA, DC, MA, WY, AZ, FL, CA, NM, NJ, TN, OH, KS, IN, NY, IL, TX, NV, NE, PA, AL, IA, GA, MO, KY, WI, OR, CO, WV, WA, ET, UT, ID, OK, MN, ME, MT, MI, NC, ND, MD, DE, VT. Quantity: 1745 Lot/code info: 00-3020: UDI-DI-07613365002737, Serial Numbers prior to 7022.002871 User/Operator Manual Versions: 41-3007EN-01 and prior
07613365002737