← Search recalls
FDA (Device enforcement)Recalled 2025-05-22class ii

Siemens Healthcare Diagnostics, Inc.: Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;

Hazard

Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.

Details from the source

Reason for recall: Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Bahamas. Quantity: 1,625 units (1450 US, 175 OUS) Lot/code info: Siemens Material Number (SMN): 11099300; UDI-DI: 00630414595771; Lot Number: 0000195412;

UPC codes

00630414595771

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.