Heraeus Medical GmbH (Dental Division): Heraeus, PALAMIX uno. Material Number: 66057893.

Hazard

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Details from the source

Reason for recall: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months). Distribution: Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Kore Quantity: 51,771 units Lot/code info: Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;

UPC codes

04260102133587

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify