FDA (Device enforcement)Recalled 2025-04-02class ii

Preat Corp: Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandi

Hazard

Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.

Details from the source

Reason for recall: Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant. Distribution: U.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA, RI, TN,TX, and WA. Quantity: 145 abutments Lot/code info: Model Number: 9001767-F UDI-DI code: 00842092166093 Lot Numbers: CN932775 ED11-202416933 ED11-20249907 ED11-202435949 CN939348 ED11-202417239 ED11-202411745 ED11-202436404 CN945176 ED11-202419976 ED11-202413957 ED11-202437437 CN946712 ED11-202422121 ED11-202414696 ED11-202437710 TEC29015668 ED11-202422577 ED11-202416838 ED11-202437908 29018601 ED11-202428254 ED11-202433998 ED11-202449612 RD10511988 ED11-202428428 ED11-202435120 242070 2905742 ED11-20249819 ED11-202435909 20-607-UCHE-000603 8-663465-000743 3-16287-002239 14-493GARIEPY-000488 20-769300-000697 20-663461ROBI-000707 30-15958-002006 7-804684-001981 13-766227-000633 6-662469CARM-000576 20-15117-001341 22-790909-001273 8-489686-000448 30-Y097-002143 19-14364-001061 10-788120-001206 7-754951-000206 3-RIEHL-001837 CO19-2024-91576 CO19-2024-68877 CO19-2024-36953 CO19-2024-24127

UPC codes

00842092166093

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify