← Search recalls
FDA (Device enforcement)Recalled 2023-05-02class ii

SenTec AG: V-Sign Sensor 2, REF VS-A/P/N, a component of Sentec Digital Monitoring System (model SDMS).

Hazard

The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode.

Details from the source

Reason for recall: The sensors may experience an out-of-box failure because after recalibration, the sensors stayed in the software mode "production" and were not reset to "released" mode. Distribution: US Nationwide distribution in the states of CA, FL, IL, MA, MS, MT, NV, and TX. Quantity: 11 in the U.S. Lot/code info: Serial numbers 338947, 338952, 338953, 338963, 338967, 338989, 338991, 338993, 338994, 338996, 338997, UDI-DI 07640121880513.

UPC codes

07640121880513

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.