← Search recalls
FDA (Device enforcement)Recalled 2023-05-22class ii

Wilson-Cook Medical Inc.: Hemospray Endoscopic Hemostat: Reference Part Numbers: a) HEMO-7, REF G56572; b) HEMO-7-EU, REF G24663;

Hazard

There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope.

Details from the source

Reason for recall: There are potential risks of the Hemospray powder adhering to the distal end of the endoscope, which can result in adhesion of the endoscope to tissue and consequent difficulty or inability to maneuver/remove the endoscope. Distribution: US Nationwide - Worldwide Distribution Quantity: 63,830 units Lot/code info: UDI/DI : a) 00827002565722, b) 00827002246638; All unexpired lot numbers

UPC codes

00827002565722, 00827002246638

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.