Boston Scientific Neuromodulation Corporation: Driver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion

Hazard

Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.

Details from the source

Reason for recall: Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury. Distribution: US Nationwide Distribution: IN, NY, IL, PA, FL, SC, TX, MA, VA, AL, WI, OK, MO, MI, OH, WV, NC, UT, NH, CA, NJ, TN, NV, AZ, NE, IA, CT, ME, LA, KY, MS, MD, DC, GA, CO, KS, ID, MN, AR, HI, NM, WA, SD, OR, DE, WY, AK, and ND Quantity: 21,533 Lot/code info: Driver UDI-DI: 00884662000574, UPN: 102-9800, All Lots. Superion Indirect Decompression System IFU (92479815-02), Superion IDS Kit IFU (92479820-02), Surgical Technique Manual (92479821-02)

UPC codes

00884662000574

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify