Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab
Reason for recall: Fails Post-Sterilization Sterile Swabs Cantilever (Bend) Testing and may be more susceptible to breakage. If the swab breaks in the nasal cavity of a patient may cause injury or medical intervention to remove part of the swab Distribution: US Nationwide distribution. Quantity: 1003 kits Lot/code info: GTIN-DI: 00850027193205 Pack Lot Number : 85438; Individual Swab Lot Number: 86445 Exp Date: 28-Oct-2023