← Search recalls
FDA (Device enforcement)Recalled 2025-05-06class ii

Olympus Corporation of the Americas: Brand Name: Powered Laser Surgical Instrument Product Name: SOLTIVE Premium SuperPulsed Laser System Model/Catalog Number: TFL-PLS, TFL-PLS-JA (Japan only); Softwa

Hazard

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

Details from the source

Reason for recall: Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents. Distribution: Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG. Quantity: 1549 units (1025 US, 524 OUS) Lot/code info: All lot/serial numbers for the following products: TFL-PLS UDI: 00821925044111; TFL-PLS-JA (Japan only);

UPC codes

00821925044111

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.