Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
Reason for recall: Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay. Distribution: Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Quantity: 4842 Lot/code info: REF/UDI-DI/Serial Numbers: HT70M-JP-NA/00839935008221/N10HT700710132, N10HT700710135, N10HT700710136, N10HT700710159, N10HT700710161, N10HT700710162, N10HT700710163, N10HT700710164, N10HT700810194, N10HT700810199, N10HT700810202, N10HT700810204, N10HT700810205, N10HT700810206, N10HT700910223, N10HT700910227, N10HT701010279, N10HT701011015, N10HT701011016, N10HT701111044, N10HT701111048, N10HT701111049, N10HT701111050, N11HT700111137, N11HT700111139, N11HT700111141, N11HT700111142, N11HT700111143, N11HT700111144, N11HT700211145, N11HT700211155, N11HT700211160, N11HT700211161, N11HT700211163, N11HT700211164, N11HT700311168, N11HT700311169, N11HT700311170, N11HT700311183, N11HT700311184, N11HT700311185, N11HT700311186, N11HT700311188, N11HT700311189, N11HT700311191, N11HT700311192, N11HT700311193, N11HT700311195, N11HT700311196, N11HT700311197, N11HT700311201, N11HT700311202, N11HT700311203, N11HT700311204, N11HT700311207, N11HT700311208, N11HT700311209, N11HT700311233, N11HT700311234, N11HT700311243, N11HT700311244, N11HT700311245, N11HT700311246, N11HT700311247, N11HT700311248, N11HT700311250, N11HT700311251, N11HT700311253, N11HT700311255, N11HT700411256, N11HT700411257, N11HT700411258, N11HT700411262, N11HT700411263, N11HT700411264, N11HT700411265, N11HT700411266, N11HT700411267, N11HT700411268, N11HT700411271, N11HT700411277, N11HT700411278, N11HT700411279, N11HT700411282, N11HT700411284, N11HT700411285, N11HT700411286, N11HT700411287, N11HT700411288, N11HT700411289, N11HT7
00839935008221, 10884521542044, 10884521542051, 10884521542082, 10884521542105, 10884521542136, 10884521542143, 10884521544178, 10884521542174, 10884521544291, 10884521544307, 10884521542181, 00839935009457, 10884521544338, 10884521542204, 10884521197855, 10884521208940, 10884521544376, 10884521542242, 10884521542259, 10884521542273, 10884521544369, 10884521542228
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.