Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.
Reason for recall: Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment. Distribution: Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX, NJ, FL, IN, MA, NY, NC, OH, LA, AZ, MO, WA, GA, UT, KY, AL, CO, IL, NE, OK, NV, SC, TN, MD, OR, AR, MS, AK, WI, DE, MT, HI, KS, ME, NH, WV, NM, CT, IA, MI, MN, ID, RI, SD, WY, ND, DC, VT, GUAM, PR and OUS (foreign) to countries of: Viet Nam, Japan, India, Australia, El Salvador, China, United Arab Emirates, Thailand, Russ Quantity: 6,166 Lot/code info: Device: REF/UDI-DI/Serial-Lot: AU680: B12185/14987666540732, B12186/14987666540688, B12187/14987666540701, B12188/14987666540725, B96694/14987666542989, B96695/14987666542996, B96696/14987666543009, C02656/14987666543306, C02657/14987666543313, N3147700/15099590346249, N3662000/14987666535301, N3662100/14987666535318, N3662200/14987666535325, N3662700/14987666535615, N3663000/14987666535646, N3910200/14987666535790, N3910400/14987666535813, N3910500/14987666535820, N3910600/14987666535912, N3910800/14987666535936, N3910900/4987666535946, N3911200/14987666535974, N3911700/14987666536575, N3911900/14987666536599, N3147100, N3147200, N3147300, N3147400, N3147600, N3148100, N3149400, N3149800. Serial: From 10436 to 2022087656. Sample Probe: MU993400/04987666522939, Lots: From 178713114 to 179433670. B66752 Kits containing MU993400, Lots: From 178713114 to 179433670
14987666540732, 14987666540688, 14987666540701, 14987666540725, 14987666542989, 14987666542996, 14987666543009, 14987666543306, 14987666543313, 15099590346249, 14987666535301, 14987666535318, 14987666535325, 14987666535615, 14987666535646, 14987666535790, 14987666535813, 14987666535820, 14987666535912, 14987666535936, 4987666535946, 14987666535974, 14987666536575, 14987666536599, 04987666522939
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.