FDA (Device enforcement)Recalled 2025-04-30class ii

Beckman Coulter Mishima K.K.: Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670

Hazard

A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.

Details from the source

Reason for recall: A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results. Distribution: Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX, NJ, FL, IN, MA, NY, NC, OH, LA, AZ, MO, WA, GA, UT, KY, AL, CO, IL, NE, OK, NV, SC, TN, MD, OR, AR, MS, AK, WI, DE, MT, HI, KS, ME, NH, WV, NM, CT, IA, MI, MN, ID, RI, SD, WY, ND, DC, VT, GUAM, PR and OUS (foreign) to countries of: Viet Nam, Japan, India, Australia, El Salvador, China, United Arab Emirates, Thailand, Russ Quantity: 10,176 Lot/code info: Device: REF/UDI-DI/Serial-Lot: AU480: B11810/14987666540664, B12183/14987666540671, B80091/16954701010015, B96692/14987666542965, B96693/14987666542972, C02654/14987666543283, C02655/14987666543290, C02845/16954701010022, C41919/14987666544174, C41920/14987666544181, N3659700/14987666535257, N3659800/14987666535264, N3659900/14987666535271, N3660000/14987666535288, N3660100/14987666535547, N3660200/14987666535554, N3660300/14987666535554, N3660400/14987666535769, N3660500/14987666536032, N3660600/14987666536049, N3660800/14987666536063, N3912400/14987666537084, N3912500/14987666537091, N3150700, N3151000, N3151300, N3151400, N3151500, N3151600, N3151900/ Serial: From 10155 to 2025021482. Sample Probe: MU993400/04987666522939, Lots: From 178713114 to 179433670. B66751 Kits containing MU993400, Lots: From 178713114 to 179433670

UPC codes

14987666540664, 14987666540671, 16954701010015, 14987666542965, 14987666542972, 14987666543283, 14987666543290, 16954701010022, 14987666544174, 14987666544181, 14987666535257, 14987666535264, 14987666535271, 14987666535288, 14987666535547, 14987666535554, 14987666535769, 14987666536032, 14987666536049, 14987666536063, 14987666537084, 14987666537091, 04987666522939

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify