← Search recalls
FDA (Device enforcement)Recalled 2024-04-23class ii

Stryker Neurovascular: Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x

Hazard

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

Details from the source

Reason for recall: Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis. Distribution: Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, Franc Quantity: 461 Lot/code info: Catalog/UDI-DI/Lot: 90182/00815742001822/116706; 90183/00815742001839/117514; 90185/00815742001853/118900; 90186/00815742001860/114054; 80051/00815742000511/47307561, 47328877, 47351811, 47398058, 47398074, 48521577, 48521585, 48540280, 48540281, 48540282, 48540283, 48540284, 48556786, 48588490, 48607589, 48676218, 48696629, 48715720, 48736116, 48736122, 48754915, 48754920, 48772012, 48772014, 48772025, 50065499, 50065511, 50065512, 50065513, 50065514, 50065515, 50065518, 50113384, 50113385, 50133176, 50133191, 50193582, 50216273, 50216274, 50216277, 50239676, 50260575, 50260583, 50280979, 50281011, 50304688, 50304712, 50304713, 50304714, 50397877, 50422572, 50422582, 50422589, 50448279, 50519274, 50519275, 50519276, 50519277, 50519285, 50519295, 50577572, 50607374, 50607376, 50607377, 50607378, 50650278, 50650279, 50650288, 50673881, 50673884, 50673908, 50673909, 50673910, 50673911, 50673930, 50694279, 50738912, 50786710, 50850697, 50869887, 50869892, 50869905, 50886487, 50912493, 50912514, 50912515, 50912516, 50912517, 50931490, 50931518, 50931519, 50950402, 50967093, 51320497, 51350503, 51350505, 51489228, 51489235, 51513311, 51513516, 51535421, 51729133; 80052/00815742000528/49021114, 49021115, 49021116, 49021117, 49021118, 49033114, 49033115, 49033116, 49033117, 49052517, 49052525, 49192125, 49192140, 49192141, 49216723, 49216724, 49216737, 53175402; 93067/00815742000672/53320201; 93068/00815742000689/48963313, 49056315, 49155966, 49216725, 49318831, 49318836, 49318837,

UPC codes

00815742001822, 00815742001839, 00815742001853, 00815742001860, 00815742000511, 00815742000528, 00815742000672, 00815742000689, 07613327126334, 07613327126358, 07613327126365

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.