Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
Reason for recall: Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision. Distribution: Worldwide - US Nationwide distribution including in the states of TX, WA, MT, GA, TN, MA, OR, ID, CA, FL, WI, UT, DE, AZ, PA, DC, LA, KY, VA, MN, NC, OH, NY, MI, CO, IN, MD, NV, AL, IL, AR, SC, OK, MO, MS, NJ, ME, Hi, KS, SD, WV, NM, AK, CT, WY, NE, ND and the countries of Canada, Argentina, Chile, Colombia, Brazil, Algeria, Austria, Belgium, Bulgaria, Croatia, Denmark, Finland, France, Germany, G Quantity: 14,977 Lot/code info: REF/UDI-DI(GTIN)/Serial Number Range/Expiration Date Range: M365DB12160/8714729985044/100209 - 753347/09-OCT-2020 through 26-MAR-2026; M365DB12320/8714729985051/100104 - 753200/09-OCT-2020 through 21-MAR-2026
8714729985044, 8714729985051
Official record: FDA (Device enforcement) ↗
Republished from the official source above; details can change — the source record controls.