FDA (Device enforcement)Recalled 2024-05-03class ii

Philips North America Llc: Ingenia 1.5T Model Number (REF): (1) 782140; (2) 782115; (3) 782101; (4) 781396; (5) 781341; (6) 781315

Hazard

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Details from the source

Reason for recall: The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire Distribution: Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom. Quantity: 2262 units Lot/code info: UDI-DI: (1) 782140 (01)00884838108646(21); (2) 782115 (01)00884838099043(21); (3) 782101 (01)00884838098275(21); (4) 781396 (01)00884838009820(21); (5) 781341 (01)00884838055322(21); (6) 781315 (01)00884838099715(21); Serial Numbers: 61019 61140 61080 61067 61091 61114 61028 61039 61040 61041 61000 61002 61003 61006 61009 61013 61017 61032 61058 61065 61066 61074 61075 61076 61097 61106 61124 61060 61001 61004 61008 61011 61023 61025 61026 61034 61035 61042 61046 61054 61055 61062 61063 61115 61078 61015 61119 61005 61020 61043 61085 61087 61088 61029 61030 61033 61036 61037 61045 61051 61052 61057 61061 61064 61068 61069 61070 61072 61095 61123 61116 61117 61125 61126 61129 61130 61027 61108 61109 61131 61144 61107 61138 61142 61143 61102 61103 61007 61014 61018 61016 61038 61073 61079 61024 61031 61059 61082 61084 61092 61093 61096 61101 61022 61044 61047 61049 61050 61053 61056 61094 61105 61112 61132 87260 70040 70229 70607 70763 84030 84108 84126 84234 84257 84596 87005 87146 70064 70113 70603 70604 70605 70704 70725 70767 70768 70794 70812 84539 70232 70279 70367 70374 70375 70461 70742 70764 70471 70780 84566 84621 84629 84667 87214 87215 70488 70832 70071 70311 70916 84098 84343 84765 70088 70110 70111 70161 70177 70204 70244 70269 70446 70448 70459 70470 70612 70662 70663 70664 70665 70666 70667 70668 70678 70691 70755 70845 70864 70866 70867 70868 70942 70943 70969 84178 84179 84180 84302 84303 84330 84381 84384 84385 84387 84451 84844 84976 87028 87040 87061 87098

UPC codes

00884838108646, 00884838099043, 00884838098275, 00884838009820, 00884838055322, 00884838099715

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify