An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid
Reason for recall: An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid Distribution: Worldwide Distribution: US (nationwide) to states of: FL IN MA MI NH NJ NY OH TX UT and OUS (Foreign) countries of: Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Brazil¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ P.R.¿China¿¿¿¿¿ Singapore¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿ Quantity: 100 units US; 328 units OUS Lot/code info: UDI-DI: (01)00630414597188(10)D389S(17)221130 (US); (01)00630414597171(10)389(17)221130 (OUS) Lot Number: 389 Exp.Date: 2022-11-30
00630414597188, 00630414597171