← Search recalls
FDA (Device enforcement)Recalled 2022-05-24class ii

Tandem Diabetes Care Inc: t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology

Is my product affected?

t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology

  • UPC 00852162004781
  • UPC 00852162004798
  • UPC 00853052007240
  • UPC 853052007257
  • UPC 853052007271
  • UPC 853052007295
  • UPC 853052007318
  • UPC 853052007967
  • UPC 853052007981
  • UPC 853052007264
  • UPC 853052007288
  • UPC 853052007301
  • UPC 853052007325
  • UPC 850006613373
  • UPC 850006613380
  • UPC 850006613700
  • UPC 850006613717
  • UPC 850006613724
  • UPC 850006613731
  • UPC 850006613229
  • UPC 850006613205
  • UPC 850006613212
  • UPC 850006613762
  • UPC 850006613779
  • UPC 850006613786
  • UPC 850006613793

Check against the official notice below — it lists the full affected range.

What happened

Tandem Diabetes Care Inc. is recalling approximately 31,100 t:slim X2 insulin pumps distributed across the US, US territories, Australia, New Zealand, and South Africa. The pumps may experience one or more of the following issues: malfunction of non-volatile memory, inaccurate or fluctuating battery life display, touchscreen remaining on, or unexpected open loop operation. These malfunctions could potentially result in hypoglycemia, hyperglycemia, or diabetic ketoacidosis. The issues can be addressed through a software update.

Affected consumers should contact Tandem Diabetes Care Inc., as the recall document does not specify what actions individual consumers should take regarding their devices.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.

Details from the source

Reason for recall: Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis. Distribution: US: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC. OUS: Australia, New Zealand, South Africa Quantity: 31,100 Lot/code info: Pump/UDI-DI/Software Versions/Distributor Instructions: t:slim X2 insulin pump/00852162004781, 00852162004798, 00853052007240/ 4.3.5.2, 4.3.6/ For Distributors, All Parts, Contact Recalling Firm to Switch Out; t:slim X2 insulin pump with Dexcom G5/853052007257, 853052007271, 853052007295, 853052007318, 853052007967, 853052007981/ 5.0.1, 5.0.4, 5.3.1.1/ For Distributors, All Parts, Contact Recalling Firm to Switch Out; t:slim X2 insulin pump with Basal-IQ technology/853052007264, 853052007288, 853052007301, 853052007325, 850006613373, 850006613380, 850006613700, 850006613717, 850006613724, 850006613731/ 6.0.4, 6.3.1.1, 6.4/ For Distributors: 1) Parts: 1000886, 1000898, 1004219, 1005706, 1006373 or 1006375, Contact Recalling Firm to Switch Out, 2) Parts: 1010004, 1010005,1010006, No Action Required, Software is Up to Date; t:slim X2 insulin pump with Control-IQ technology/ 850006613229, 850006613205, 850006613212, 850006613762, 850006613779, 850006613786, 850006613793/ 7.3.1.1, 7.4/ For Distributors: 1) Parts: 1005011, 1005012, 1006402 or 1006404, Contact Recalling Firm to Switch Out, 2) Parts: 1010009,1010010, 1010011, No Action Required, Software is Up to Date

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.