← Search recalls
FDA (Device enforcement)Recalled 2026-03-19class i

B Braun Medical Inc: B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096

Hazard

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Details from the source

Reason for recall: Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. Distribution: Worldwide distribution - US Nationwide and the country of Canada. Quantity: 2,653,711 units Lot/code info: Model Number: SL-2010M2096; UDI-DI Primary: 04046964367786; UDI-DI Unit: 04046964367779; All lots manufactured since 25MAY2025.

UPC codes

04046964367786, 04046964367779

Source

Official record: FDA (Device enforcement)

Republished from the official source above; details can change — the source record controls.

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.