What happened

Fresenius Medical Care is recalling 6 units of the AquaC UNO H reverse osmosis unit (Fresenius Part Number 24-0970-1), a device used with hemodialysis systems. The recall affects units distributed nationwide in Arizona, California, the District of Columbia, Florida, Georgia, Indiana, Massachusetts, Nevada, New York, Pennsylvania, and Texas. The devices are being recalled because internal ethernet cables are missing one or both ferrite cores, which means the electromagnetic compatibility of the device cannot be ensured.

The specific serial numbers involved in this recall are 0CHS3987, 0CHS4108, 1CHS5372, 1CHS5405, 9CHS2881, and 9CHS3090. The source document does not specify what actions affected consumers should take in response to this recall.

Written by software from the official notice below. Not reviewed by a lawyer.

Hazard

Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

Details from the source

Reason for recall: Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured Distribution: US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX. Quantity: 6 units Lot/code info: UDI: (01)4251285200234(11)200422(21)0CHS3987 (01)4251285200234(11)200526(21)0CHS4108 (01)4251285200234(11)210623(21)1CHS5372 (01)4251285200234(11)210707(21)1CHS5405 (01)4251285200234(11)190305(21)9CHS2881 (01)4251285200234(11)190529(21)9CHS3090 Serial Numbers: 0CHS3987 0CHS4108 1CHS5372 1CHS5405 9CHS2881 9CHS3090

Source

Official record: FDA (Device enforcement) ↗

Republished from the official source above; details can change — the source record controls.

Related recalls

All FDA (Device enforcement) recalls

Hurt by this product? Start a free Case Check

ProductRisk is not a law firm and does not provide legal services or legal advice. ProductRisk is a free consumer product-safety information service operated by Plaintify, Inc., a technology company. If you ask to be connected with a law firm, any legal services are provided by that firm — not by ProductRisk or Plaintify.

Privacy · Plaintify